Test Code LAB114 POTASSIUM
Additional Codes
Internal: 10386
External: K
Beaker: LAB114
PERFORMED AT
SBH LABORATORY
SRH LABORATORY
SMH LABORATORY
SPH LABORATORY
HDH LABORATORY
Specimen Requirements
Container Type: Lithium Heparin (Green Top) or Serum Separator (Gold Top)
Specimen Type: Plasma or Serum
Preferred Volume: 1 mL
Emergency Minimum Volume: 0.3 mL
Specimen Processing: Separate serum or plasma from cells as soon as possible by centrifugation. Store and transport refrigerated.
Unacceptable Conditions: Hemolyzed specimens. EDTA Plasma (Lavender Top).
Stability:
Temperature | Time |
---|---|
Room Temp | 2 weeks |
Refrigerated | 2 weeks |
Frozen (-20 øC) | 2 weeks |
Method
ISE
Synonyms
K+
CPT Codes
84132
Department
Chemistry
Test Schedule
Daily & STAT
Reference Lab
SCHS
Remarks
Reference ranges differ between serum and plasma specimens due to clot formation in serum.