Test Code LAB5157 COVID-19 Ag Test
Additional Codes
Internal / External / Beaker: LAB5157
Specimen Requirements
Acceptable specimen: Dry nasal swab returned in sterile container (dry transport tube)
Sample Transport and Storage: Samples should be tested as soon as possible after collection. Nasal swabs are stable for up to 120-hours (5-days) at room temperature or 2° to 8°C in a clean, dry transport tube.
Method
Lateral flow immunoassay
Synonyms
COVID Antigen test, COVID Ag, COVID Antigen, SARS Antigen test, SARS Ag
Test Schedule
Test is performed 24 hours a day seven days a week.
Performed At
SBH Laboratory
Department
Bend Serology Department
CPT Code
87811
Specimen Collection
Prior to collecting the nasal swab, the patient should be instructed to blow their nose. To collect a nasal swab sample, insert the entire absorbent tip of the swab usually ½ to ¾ of an inch (1 to 1.5 cm) inside the nostril and firmly sample the nasal wall by rotating the swab in a circular path against the nasal wall at least 4 times. Take approximately 15 seconds to collect the sample. Be sure to collect any nasal drainage that may be present on the swab. Sample both nostrils with same swab