Test Code TTSC Thrombin Time (Bovine), Plasma
Additional Codes
Internal: 11395
External: TT
Beaker: LAB324
Specimen Required
Specimen Type: Platelet-poor plasma
Patient Preparation: Fasting preferred
Collection Container/Tube: Light-blue top (3.2% sodium citrate)
Submission Container/Tube: Plastic vial
Specimen Volume: 1 mL
Collection Instructions:
1. For complete instructions, see Coagulation Guidelines for Specimen Handling and Processing.
2. Centrifuge, transfer all plasma into a plastic vial, and centrifuge plasma again.
3. Aliquot plasma into a separate plastic vial leaving 0.25 mL in the bottom of centrifuged vial.
4. Freeze plasma immediately (no longer than 4 hours after collection) at -20° C or, ideally, -40° C or below.
Additional Information:
1. Double-centrifuged specimen is critical for accurate results as platelet contamination may cause spurious results.
2. Each coagulation assay requested should have its own vial.
Forms
If not ordering electronically, complete, print, and send a Coagulation Test Request (T753) with the specimen.
Secondary ID
602184Useful For
Detecting or excluding the presence of heparin or heparin-like anticoagulants (which act by enhancing antithrombin's inhibition of thrombin and other procoagulant enzymes) when used in conjunction with the reptilase time (RT) in evaluating unexplained prolonged clotting times
Identifying the cause of a prolonged prothrombin time, activated partial thromboplastin time, or dilute Russell viper venom time when used in conjunction with the RT and fibrinogen assay
Special Instructions
Method Name
Optical Clot-Based
Reporting Name
Thrombin Time (Bovine), PSpecimen Type
Plasma Na CitSpecimen Minimum Volume
0.5 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Plasma Na Cit | Frozen | 14 days |
Reject Due To
Gross hemolysis | Reject |
Gross lipemia | Reject |
Gross icterus | Reject |
Reference Values
15.8-24.9 seconds
Day(s) Performed
Monday through Saturday
Report Available
1 to 4 daysPerforming Laboratory
Mayo Clinic Laboratories in RochesterTest Classification
This test has been cleared, approved, or is exempt by the US Food and Drug Administration and is used per manufacturer's instructions. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.CPT Code Information
85670
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
TTSC | Thrombin Time (Bovine), P | 46717-5 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
TTSC | Thrombin Time (Bovine), P | 46717-5 |